ZY Series Oxygen Concentrator Valve Solutions

Integrated 4-Way 5-Port Valve Assembly (NO) | Micro-Direct Acting Pilot Valve (NC) | Medical-Grade PSA Components | 1.2W Low Power.

  • Standard Main Valve: ZY01-562-CD (NO) – The industry-standard 4-Way 5-Port assembly for 3L-10L oxygen concentrators.

  • Auxiliary Control: ZY01-372-1CD (NC) – Compact pilot unit for pressure balancing, rapid exhaust, or bypass logic.

  • Zero-Pressure Startup: Reliable switching from 0 Bar to 0.7 MPa (0 – 101.5 Psi).

  • High-Speed Response: Optimized for high-frequency PSA cycling (2700/min) with minimal latency.

SKU: SV-ZY01 Categories: , Tag:

The ZY Series offers precision pneumatic control for Medical Oxygen Generation. The Integrated NO Assembly is designed as the primary switching hub for molecular sieve beds, ensuring fail-safe venting during power loss. The NC Micro-Valve serves as a versatile component for engineers designing custom manifolds or requiring precise shut-off control for auxiliary gas paths. Both units feature 1.2W ultra-low power coils to prevent overheating during continuous 24/7 medical use.

Technical Specifications

  • Assembly Type: 4-Way 5-Port (Normally Open)

  • Component Type: Micro Direct-Acting (Normally Closed)

  • Operation: Direct Acting
  • Orifice Size: 12mm

  • Working Pressure: 0.05 – 0.3 MPa (7.25 – 43.5 Psi)

  • Maximum Pressure: Up to 0.7 MPa (101.5 Psi)

  • Working Temperature: -10°C to 60°C (14°F to 140°F)

  • Voltage Options: 6V / 12V / 24V DC (Customizable)

  • Power Consumption: 1.2W – 1.3W

  • Media: Compressed Air / Oxygen

  • Certification: CE